Calibration of Littleton measuring instruments using test specimens of reference glass
Date
Authors
Editor
Advisor
Volume
Issue
Journal
Series Titel
Book Title
Publisher
Supplementary Material
Other Versions
Link to publishers' Version
Abstract
The determination of the softening point of glass according to Littleton (Littleton point) is carried out by a conventional method of measurement. The measuring device and the method of measurement are standardized. The calibration of Littleton measuring instruments is performed with the aid of test specimens manufactured in the laboratories of the glass industry from standard glass with a known Littleton point using various technologies. Owing to the availability of test specimens which are machine-made and traced back to a standard measuring device and whose Littleton point is certified, deficiencies of the usual calibration procedure are avoided and the comparability of the measurement results is improved.
Die Bestimmung des Erweichungspunktes von Glas nach Littleton (Littleton-Punkt) wird nach den konventionellen Meßverfahren durchgeführt. Meßeinrichtung und Meßverfahren sind genormt. Die Kalibrierung der Littleton-Meßgeräte wird mit Hilfe von Prüfkörpern durchgeführt, die in den Laboratorien der Glasindustrie unter Anwendung verschiedener Techniken aus Standardglas mit einem bekannten Littleton-Punkt hergestellt werden. Auf Grund der Verfügbarkeit von maschinell gefertigten Prüfkörpern, die auf eine Normalmeßeinrichtung zurückgeführt sind und deren Littleton-Punkt zertifiziert ist, werden Mängel der herkömmlichen Kalibrierungsmethode vermieden und die Vergleichbarkeit der Meßergebnisse verbessert.