Test beam measurement of the first prototype of the fast silicon pixel monolithic detector for the TT-PET project
| dc.bibliographicCitation.firstPage | P04015 | |
| dc.bibliographicCitation.issue | 4 | |
| dc.bibliographicCitation.journalTitle | Journal of Instrumentation | eng |
| dc.bibliographicCitation.volume | 13 | |
| dc.contributor.author | Paolozzi, L. | |
| dc.contributor.author | Bandi, Y. | |
| dc.contributor.author | Benoit, M. | |
| dc.contributor.author | Cardarelli, R. | |
| dc.contributor.author | Débieux, S. | |
| dc.contributor.author | Forshaw, D. | |
| dc.contributor.author | Hayakawa, D. | |
| dc.contributor.author | Iacobucci, G. | |
| dc.contributor.author | Kaynak, M. | |
| dc.contributor.author | Miucci, A. | |
| dc.contributor.author | Nessi, M. | |
| dc.contributor.author | Ratib, O. | |
| dc.contributor.author | Ripiccini, E. | |
| dc.contributor.author | Rücker, H. | |
| dc.contributor.author | Valerio, P. | |
| dc.contributor.author | Weber, M. | |
| dc.date.accessioned | 2023-01-24T08:05:52Z | |
| dc.date.available | 2023-01-24T08:05:52Z | |
| dc.date.issued | 2018 | |
| dc.description.abstract | The TT-PET collaboration is developing a PET scanner for small animals with 30 ps time-of-flight resolution and sub-millimetre 3D detection granularity. The sensitive element of the scanner is a monolithic silicon pixel detector based on state-of-the-art SiGe BiCMOS technology. The first ASIC prototype for the TT-PET was produced and tested in the laboratory and with minimum ionizing particles. The electronics exhibit an equivalent noise charge below 600 e− RMS and a pulse rise time of less than 2 ns , in accordance with the simulations. The pixels with a capacitance of 0.8 pF were measured to have a detection efficiency greater than 99% and, although in the absence of the post-processing, a time resolution of approximately 200 ps . | eng |
| dc.description.version | publishedVersion | eng |
| dc.identifier.uri | https://oa.tib.eu/renate/handle/123456789/10984 | |
| dc.identifier.uri | http://dx.doi.org/10.34657/10010 | |
| dc.language.iso | eng | |
| dc.publisher | London : Inst. of Physics | |
| dc.relation.doi | https://doi.org/10.1088/1748-0221/13/04/p04015 | |
| dc.relation.essn | 1748-0221 | |
| dc.rights.license | CC BY 3.0 Unported | |
| dc.rights.uri | https://creativecommons.org/licenses/by/3.0 | |
| dc.subject.ddc | 610 | |
| dc.subject.other | Front-end electronics for detector readout | eng |
| dc.subject.other | Gamma camera SPECT PET PET/CT coronary CT angiography (CTA) | eng |
| dc.subject.other | Si microstrip and pad detectors | eng |
| dc.subject.other | Timing detectors | eng |
| dc.title | Test beam measurement of the first prototype of the fast silicon pixel monolithic detector for the TT-PET project | eng |
| dc.type | Article | eng |
| dc.type | Text | eng |
| tib.accessRights | openAccess | |
| wgl.contributor | IHP | |
| wgl.subject | Medizin, Gesundheit | ger |
| wgl.type | Zeitschriftenartikel | ger |
Files
Original bundle
1 - 1 of 1
Loading...
- Name:
- Test_beam_measurement.pdf
- Size:
- 2.6 MB
- Format:
- Adobe Portable Document Format
- Description: