dc.description.abstractNewly developed materials for blood-contacting devices need to undergo hemocompatibility testing to prove compliance with clinical requirements. However, many current in vitro models disregard the influence of flow conditions and blood exchange as it occurs in vivo. Here, we present a flow model which allows testing of blood-surface interactions under more physiological conditions. This modular platform consists of a triple-pump-chip and a microchannel-chip with a customizable surface. Flow conditions can be adjusted individually within the physiological range. A performance test with whole blood confirmed the hemocompatibility of our modular platform. Hemolysis was negligible, inflammation and hemostasis parameters were comparable to those detected in a previously established quasi-static whole blood screening chamber. The steady supply of fresh blood avoids secondary effects by nonphysiological accumulation of activation products. Experiments with three subsequently tested biomaterials showed results similar to literature and our own experience. The reported results suggest that our developed flow model allows the evaluation of blood-contacting materials under physiological flow conditions. By adjusting the occurring wall shear stress, the model can be adapted for selected test conditions.eng
dc.description.versionpublishedVersioneng
dc.identifier.urihttps://oa.tib.eu/renate/handle/123456789/12053
dc.identifier.urihttp://dx.doi.org/10.34657/11086
dc.language.isoeng
dc.publisherBerlin : De Gruyter
dc.relation.doihttps://doi.org/10.1515/cdbme-2021-2044
dc.relation.essn2364-5504
dc.rights.licenseCC BY 4.0 Unported
dc.rights.urihttps://creativecommons.org/licenses/by/4.0
dc.subject.ddc570
dc.subject.otherbloodsurface interactionseng
dc.subject.otherhemocompatibilityeng
dc.subject.otherin vitro flow modeleng
dc.titleA modular in vitro flow model to analyse blood-surface interactions under physiological conditionseng
dc.typeArticleeng
dc.typeTexteng
tib.accessRightsopenAccess
wgl.contributorIPF
wgl.subjectBiowissenschaften/Biologieger
wgl.typeZeitschriftenartikelger
Files
Original bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
A_modular_in_vitro_flow_model.pdf
Size:
1.57 MB
Format:
Adobe Portable Document Format
Description: