Test beam measurement of the first prototype of the fast silicon pixel monolithic detector for the TT-PET project

2018, Paolozzi, L., Bandi, Y., Benoit, M., Cardarelli, R., Débieux, S., Forshaw, D., Hayakawa, D., Iacobucci, G., Kaynak, M., Miucci, A., Nessi, M., Ratib, O., Ripiccini, E., Rücker, H., Valerio, P., Weber, M.

The TT-PET collaboration is developing a PET scanner for small animals with 30 ps time-of-flight resolution and sub-millimetre 3D detection granularity. The sensitive element of the scanner is a monolithic silicon pixel detector based on state-of-the-art SiGe BiCMOS technology. The first ASIC prototype for the TT-PET was produced and tested in the laboratory and with minimum ionizing particles. The electronics exhibit an equivalent noise charge below 600 e− RMS and a pulse rise time of less than 2 ns , in accordance with the simulations. The pixels with a capacitance of 0.8 pF were measured to have a detection efficiency greater than 99% and, although in the absence of the post-processing, a time resolution of approximately 200 ps .