Design and Fabrication of a BiCMOS Dielectric Sensor for Viscosity Measurements: A Possible Solution for Early Detection of COPD

Abstract

The viscosity variation of sputum is a common symptom of the progression of Chronic Obstructive Pulmonary Disease (COPD). Since the hydration of the sputum defines its viscosity level, dielectric sensors could be used for the characterization of sputum samples collected from patients for early diagnosis of COPD. In this work, a CMOS-based dielectric sensor for the real-time monitoring of sputum viscosity was designed and fabricated. A proper packaging for the ESD-protection and short-circuit prevention of the sensor was developed. The performance evaluation results show that the radio frequency sensor is capable of measuring dielectric constant of biofluids with an accuracy of 4.17%. Integration of this sensor into a portable system will result in a hand-held device capable of measuring viscosity of sputum samples of COPD-patients for diagnostic purposes.

Description
Keywords
Biosensors, CMOS viscosity sensor, Dielectric measurements, Precision diagnostics, Radio frequency sensor, Sputum–Mucin, Viscometer
Citation
Soltani Zarrin, P., Jamal, F. I., Guha, S., Wessel, J., Kissinger, D., & Wenger, C. (2018). Design and Fabrication of a BiCMOS Dielectric Sensor for Viscosity Measurements: A Possible Solution for Early Detection of COPD. 8(3). https://doi.org//10.3390/bios8030078
License
CC BY 4.0 Unported