The Structure, Properties and Application of Solid Contact Glass Mini-Electrode
dc.bibliographicCitation.firstPage | 332 | |
dc.bibliographicCitation.journalTitle | Glass Science and Technology | |
dc.bibliographicCitation.lastPage | 337 | |
dc.bibliographicCitation.volume | 77c | |
dc.contributor.author | Wang, C. | |
dc.contributor.author | Chen, M. | |
dc.contributor.author | Dong, S. | |
dc.contributor.author | Tao, Y. | |
dc.contributor.author | Qi, J. | |
dc.contributor.author | Lepnev, G. | |
dc.date.accessioned | 2024-08-23T15:03:38Z | |
dc.date.available | 2024-08-23T15:03:38Z | |
dc.date.issued | 2004 | |
dc.description.abstract | The glass mini-electrode, with solid (Li/Pb alloy) contact with lithium silicate glass (containing La₂O₃ and ZrO₂), is prepared, which diameter is 2 mm and length 7mm. The measuring range of pH value is 1-9; the transition parameter is larger than 90%; response time is less than 10s. The stomach juices of 18 white rats are surveyed. The results show that pH values of them are within 2-5 and the average deviation is less than 0.1 compared to that measured by PB-20 Sartorius acidimeter. There is a great potential for the mini-electrode to be applied to examining mortal stomach juice, for diagnosing stomach- disorder and cancer in advance. | eng |
dc.description.version | publishedVersion | |
dc.identifier.uri | https://oa.tib.eu/renate/handle/123456789/14923 | |
dc.identifier.uri | https://doi.org/10.34657/13945 | |
dc.language.iso | eng | |
dc.publisher | Offenbach : Verlag der Deutschen Glastechnischen Gesellschaft | |
dc.relation.issn | 0946-7475 | |
dc.rights.license | CC BY 3.0 DE | |
dc.rights.uri | https://creativecommons.org/licenses/by/3.0/de/ | |
dc.subject.ddc | 660 | |
dc.subject.gnd | Konferenzschrift | |
dc.title | The Structure, Properties and Application of Solid Contact Glass Mini-Electrode | eng |
dc.type | Article | |
dc.type | Text | |
tib.accessRights | openAccess |