Modified UV-VIS-NIR spectrophotometric investigation method for unprepared glass samples

dc.bibliographicCitation.firstPage161
dc.bibliographicCitation.lastPage167
dc.bibliographicCitation.volume64
dc.contributor.authorBertolin, Mario
dc.date.accessioned2024-08-27T15:47:00Z
dc.date.available2024-08-27T15:47:00Z
dc.date.issued1991
dc.description.abstractThe paper reports about practical experience in industrial glass production control by a slightly modified ultraviolet and visible near infrared (UV-VIS-NIR) spectrophotometric method. A key element for the successful implementation of the method has been identified in the photodiode used as detector in the commercially available spectrophotometers. With some software tools and a specific sample holder it is possible to implement an accurate measurement of spectral features an a commercial glass sample without significant physical preparation. The glass making process can be kept under tight control without the need of laboratory technicans. The paper shows that by the carefully chosen combination of detector and data-handling tools glass quality determining factors (e.g. redox equlibria) can be measured quickly and accurately for quality assurance of the final product. In the case of flint glasses it is possible to calculate the concentrations of key polyvalent elements (Fe2+ , Fe3+ , Cr3+) as well as derived parameters of technological relevance.
dc.description.versionpublishedVersion
dc.identifier.urihttps://oa.tib.eu/renate/handle/123456789/15064
dc.identifier.urihttps://doi.org/10.34657/14086
dc.language.isoeng
dc.publisherOffenbach : Verlag der Deutschen Glastechnischen Gesellschaft
dc.relation.ispartofseriesGlastechnische Berichte
dc.relation.issn0946-7475
dc.rights.licenseCC BY 3.0 DE
dc.rights.urihttps://creativecommons.org/licenses/by/3.0/de/
dc.subject.ddc660
dc.titleModified UV-VIS-NIR spectrophotometric investigation method for unprepared glass samples
dc.typearticle