The 1-Megapixel pnCCD detector for the Small Quantum Systems Instrument at the European XFEL: system and operation aspects

dc.bibliographicCitation.firstPage576eng
dc.bibliographicCitation.issue2eng
dc.bibliographicCitation.lastPage587eng
dc.bibliographicCitation.volume28eng
dc.contributor.authorKuster, Markus
dc.contributor.authorAhmed, Karim
dc.contributor.authorBallak, Kai Erik
dc.contributor.authorDanilevski, Cyril
dc.contributor.authorEkmedžić, Marko
dc.contributor.authorFernandes, Bruno
dc.contributor.authorGessler, Patrick
dc.contributor.authorHartmann, Robert
dc.contributor.authorHauf, Steffen
dc.contributor.authorHoll, Peter
dc.contributor.authorMeyer, Michael
dc.contributor.authorMontaño, Jacobo
dc.contributor.authorMünnich, Astrid
dc.contributor.authorOvcharenko, Yevheniy
dc.contributor.authorRennhack, Nils
dc.contributor.authorRüter, Tonn
dc.contributor.authorRupp, Daniela
dc.contributor.authorSchlosser, Dieter
dc.contributor.authorSetoodehnia, Kiana
dc.contributor.authorSchmitt, Rüdiger
dc.contributor.authorStrüder, Lothar
dc.contributor.authorTanyag, Rico Mayro P.
dc.contributor.authorUlmer, Anatoli
dc.contributor.authorYousef, Hazem
dc.date.accessioned2022-03-24T12:49:21Z
dc.date.available2022-03-24T12:49:21Z
dc.date.issued2021
dc.description.abstractThe X-ray free-electron lasers that became available during the last decade, like the European XFEL (EuXFEL), place high demands on their instrumentation. Especially at low photon energies below 1 keV, detectors with high sensitivity, and consequently low noise and high quantum efficiency, are required to enable facility users to fully exploit the scientific potential of the photon source. A 1-Megapixel pnCCD detector with a 1024 × 1024 pixel format has been installed and commissioned for imaging applications at the Nano-Sized Quantum System (NQS) station of the Small Quantum System (SQS) instrument at EuXFEL. The instrument is currently operating in the energy range between 0.5 and 3 keV and the NQS station is designed for investigations of the interaction of intense FEL pulses with clusters, nano-particles and small bio-molecules, by combining photo-ion and photo-electron spectroscopy with coherent diffraction imaging techniques. The core of the imaging detector is a pn-type charge coupled device (pnCCD) with a pixel pitch of 75 µm × 75 µm. Depending on the experimental scenario, the pnCCD enables imaging of single photons thanks to its very low electronic noise of 3 e− and high quantum efficiency. Here an overview on the EuXFEL pnCCD detector and the results from the commissioning and first user operation at the SQS experiment in June 2019 are presented. The detailed descriptions of the detector design and capabilities, its implementation at EuXFEL both mechanically and from the controls side as well as important data correction steps aim to provide useful background for users planning and analyzing experiments at EuXFEL and may serve as a benchmark for comparing and planning future endstations at other FELs.eng
dc.description.versionpublishedVersioneng
dc.identifier.urihttps://oa.tib.eu/renate/handle/123456789/8358
dc.identifier.urihttps://doi.org/10.34657/7396
dc.language.isoengeng
dc.publisherChester : IUCreng
dc.relation.doihttps://doi.org/10.1107/S1600577520015659
dc.relation.essn1600-5775
dc.relation.ispartofseriesJournal of synchrotron radiation 28 (2021), Nr. 2eng
dc.rights.licenseCC BY 4.0 Unportedeng
dc.rights.urihttps://creativecommons.org/licenses/by/4.0/eng
dc.subjectX-ray CCDeng
dc.subjectfree-electron lasereng
dc.subjectphoton detectioneng
dc.subjectpnCCDeng
dc.subjectX-ray CCDeng
dc.subject.ddc540eng
dc.subject.ddc550eng
dc.titleThe 1-Megapixel pnCCD detector for the Small Quantum Systems Instrument at the European XFEL: system and operation aspectseng
dc.typearticleeng
dc.typeTexteng
dcterms.bibliographicCitation.journalTitleJournal of synchrotron radiationeng
tib.accessRightsopenAccesseng
wgl.contributorMBIeng
wgl.subjectChemieeng
wgl.typeZeitschriftenartikeleng
Files
Original bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
The_1-Megapixel_pnCCD_detector.pdf
Size:
1.8 MB
Format:
Adobe Portable Document Format
Description:
Collections