A monolithic ASIC demonstrator for the Thin Time-of-Flight PET scanner

Abstract

Time-of-flight measurement is an important advancement in PET scanners to improve image reconstruction with a lower delivered radiation dose. This article describes the monolithic ASIC for the TT-PET project, a novel idea for a high-precision PET scanner for small animals. The chip uses a SiGe Bi-CMOS process for timing measurements, integrating a fully-depleted pixel matrix with a low-power BJT-based front-end per channel, integrated on the same 100 µm thick die. The target timing resolution of the scanner is 30 ps RMS for electrons from the conversion of 511 keV photons. The system will include 1.6 million channels across almost 2000 different chips. A full-featured demonstrator chip with a 3×10 matrix of 500×500 µm2 pixels was fabricated to validate each block. Its design and experimental results are presented here. © 2019 CERN.

Description
Keywords
Analogue electronic circuits, coronary CT angiography (CTA), Front-end electronics for detector readout, Gamma camera, PET PET/CT, SPECT, Timing detectors
Citation
Valerio, P., Cardarelli, R., Iacobucci, G., Paolozzi, L., Ripiccini, E., Hayakawa, D., et al. (2019). A monolithic ASIC demonstrator for the Thin Time-of-Flight PET scanner. 14(7). https://doi.org//10.1088/1748-0221/14/07/P07013