The use of matrix-specific calibrations for oxygen in analytical glow discharge spectrometry

Loading...
Thumbnail Image

Date

Editor

Advisor

Volume

406

Issue

29

Journal

Analytical and Bioanalytical Chemistry

Series Titel

Book Title

Publisher

Dordrecht : Springer

Supplementary Material

Other Versions

Abstract

The performance of glow discharge optical emission spectroscopy and mass spectrometry for oxygen determination is investigated using a set of new conductive samples containing oxygen in the percent range in three different matrices (Al, Mg, and Cu) prepared by a sintering process. The sputtering rate corrected calibrations obtained at standard conditions for the 4 mm anode (700 V, 20 mA) in GD-OES are matrix independent for Mg and Al but not for Cu. The importance of a "blue shifted" line of oxygen at 130.22 nm (first reported by Köster) for quantitative analyses by GD-OES is confirmed. Matrix-specific calibrations for oxygen in GD-MS are presented. Two source concepts - fast flow (ELEMENT GD) and low gas flow (VG9000) - are evaluated obtaining higher sensitivity with the static flow source. Additional experiments using Ar-He mixtures or μs pulsed GD are carried out in ELEMENT GD aiming to improve the oxygen sensitivity.

Description

Keywords GND

Conference

Publication Type

Article

Version

publishedVersion

Collections

License

CC BY 4.0 Unported