simpleISM—A straight forward guide to upgrade from confocal to ISM

Loading...
Thumbnail Image
Date
2022
Volume
17
Issue
12
Journal
Series Titel
Book Title
Publisher
San Francisco, California, US : PLOS
Abstract

Resolution in a confocal laser scanning microscopes (CLSM) can be improved if the pinhole is closed. But closing the pinhole will deteriorate the signal to noise ratio (SNR). A simple technique to improve the SNR while keeping the resolution same by upgrading the system to an image scanning microscope. In this paper, we explain in detail, based on an Olympus Fluoview 300 system, how a scanning microscope can be upgraded into an image scanning microscope (ISM) using a simple camera-based detector and an Arduino Due providing a galvo driving and camera synchronization signals. We could confirm a resolution improvement as well as superconcentration and made the interesting observation of a reduced influence of laser fluctuations.

Description
Keywords
animal cell, BPAEC cell line, calibration, confocal laser scanning microscopy, controlled study, image reconstruction, nonhuman, optical resolution, point spread function, scanning electron microscopy, signal noise ratio, confocal microscopy, laser, procedures
Citation
Goswami, M., Lachmann, R., Kretschmer, R., & Heintzmann, R. (2022). simpleISM—A straight forward guide to upgrade from confocal to ISM. 17(12). https://doi.org//10.1371/journal.pone.0279378
Collections
License
CC BY 4.0 Unported